|
THE CHARACTERIZATION OF PARTICULATE DEBRIS OBTAINED FROM FAILED
ORTHOPEDIC IMPLANTS: | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |
Metals at the top of the table form oxides quite readily while
those at the bottom form passive layers much more slowly (or,
in the case of gold, not at all). The tenacity of the passive
layer is also higher for the metals at the top of the table. However,
in the presence of mechanical abrasion, even the most tenacious
passive film can be breached. Once this occurs, the surrounding
chemistry (i.e., the availability of free oxygen) and the
nature of the metal-oxide bond will determine the rate at which
the protective layer is repaired. If an insufficient amount of
oxygen is available, the layer may remain damaged and corrosion
is likely to occur. A more detailed discussion of the behavior
of metals in a corrosive environment will be presented in the
later section on The Service Environment for the Implant.
Titanium was discovered in 1790 and is the earth's ninth most
abundant element. The Kroll process (developed in 1936) allowed
titanium to be produced in commercial quantities. The Kroll process
involves the chlorination of the raw ore to produce titanium tetrachloride
(TiCl4). This chloride compound
is then reduced with solid magnesium metal in an inert atmosphere
to produce MgCl2 and a porous ingot
of titanium (known as a sponge). Iron chloride and residual magnesium
chloride are then leached out to purify the ingot, which is then
densified. The titanium used in the manufacture of modular implant
components is of a nominal purity of 99.0% ASTM Grade 4 with alloy
additions as specified in Table 6.
| Chemical Composition, wt% | |||||
|---|---|---|---|---|---|
| C | H | O | N | Fe | Other |
| 0.10 | 0.0100 | 0.40 | 0.05 | 0.30 | ----- |
Titanium is an allotropic element. At room temperature, the hexagonal
close packed (hcp) structure known as the a
phase is thermodynamically stable. When heated to temperatures
over 883°C (1621°F), it transforms into the body centered
cubic (bcc) b phase. In the annealed
condition, it exhibits the mechanical properties given in Table
7. Commercially pure (CP) titanium is often used in medical
devices for its relatively easy formability (compared to titanium
alloys) and high biocompatibilty. In the case of the stems investigated
for this study, the porous surfaces are fabricated from a CP titanium
wire mesh pad. The titanium wires are chopped, kinked, and then
diffusion bonded into the form of a pad approximately 2 mm thick.
The pads are then cut to their final shape to fit recessions on
the side and edge of the implant stem. The pads are then hot isostatically
pressed (HIP'ed) and diffusion bonded onto the stem surface. Diffusion
bonding is preferred over sintering because the temperatures required
to sinter porous surfaces onto the stem base material result in
very extensive grain growth in the stem, with grains often attaining
average volumes of 1 cm3. These large grained
components are likely to exhibit insufficient strength when subjected
to large or sudden loads.
| Density | 4.51 g/cm3 (~60% of steel) |
| Minimum Yield Stress | 480 MPa |
| Minimum Ultimate Tensile Strength | 550 MPa |
| Young's Modulus of Elasticity | 102.7 GPa |
| Poisson's Ratio | 0.34 |
| Hardness | 265 (Brinell), 300 Knoop |
| Coefficient of Thermal Expansion | 8.64 x 10-6/°C |
| Solidus/Liquidus | 1725°C |
| Melting Point | 1668±10°C |
| Specific Heat (25°C) | 0.518 J/kg K |
Ti6Al4V/a + b
Alloys
In the 1950's, strengthenable titanium alloys were sought in order
to answer a need for materials with high specific strength, low
creep, high melting temperature, and high corrosion resistance
for aircraft components such as jet engines. Although considered
in the early 1950s, alloys of titanium were not used for surgical
applications until the 1960's. Now, titanium alloys such as Ti-6Al-4V
are beginning to rival stainless steel as well as CoCrMo alloys
in terms of implantation uses.
As shown by the Ti-Al phase diagram in Figure 2, the addition
of 6 wt% of aluminum to titanium serves to increase the temperature
at which the a phase is stable. The
a phase is desirable for good strength
and toughness and resistance to oxygen contamination at elevated
forming temperatures to which this particular type of implant
is exposed. Aluminum exhibits a strong solution strengthening
effect when added to a titanium matrix, as shown in Figure
3.



The addition of vanadium serves to stabilize the b
phase. An addition of only 4 at% of vanadium reduces the
a-b transus temperature nearly 200°C, as seen in the
Ti-V phase diagram in Figure 4. The bcc b
phase allows for better formability at lower temperatures, but
is more susceptible to atmospheric contamination The a
phase is much more resistant to oxygen contamination and exhibits
good strength and toughness. When the a
phase is in the platelet Widmänstatten form, it provides
excellent fatigue crack arresting properties. A stabilized b
phase can be retained upon quenching. In the case of the Ti6Al4V
alloy, the b phase can even be retained
after a furnace cool. The micrograph presented in Figure 5 shows
an a+b Ti6Al4V alloy that was forged
and then aged at an intermediate temperature with a subsequent
furnace cool to room temperature.

Zirconium and hafnium have extensive solid solubilities in both
the alpha and beta phases of titanium. They do not strongly promote
phase stability but they retard the rates of transformation and
are useful as strengthening agents. Oxygen, nitrogen, and carbon
are alpha stabilizers that raise the transformation temperature.
As discussed in the next section, hydrogen is a beta stabilizer
and lowers the transformation temperature. Niobium is beta isomorphous
(bcc) and will not form intermetallic compounds with titanium.
The Ti-13Zr-13Nb alloy, mentioned above as a low modulus substitute
for Ti6Al4V (Section 3, Overview of the Materials Science of
Orthopedic Implant Systems) takes advantage of the solid solubility
of these two alloying elements (Nb and Zr) to produce a compound
of lower modulus, most likely by 'over-saturating' the titanium
matrix beyond the point of maximum solid solution strengthening.
Effect of Hydrogen on Microstructure
There are many potential sources for hydrogen to be introduced
into the implant during its manufacture and service. Residual
hydrogen in the ingot is unavoidable. However, improved refining
and furnace techniques have reduced the amount of hydrogen left
in the metal after melting. Great care must be taken during the
forging operations and subsequent handling prior to an oxidation
treatment such that no hydrocarbons (gaseous or liquid) such as
methane, or oils from machinery (or even bare hands!) are allowed
to contact the bare surface of the component. Though once considered
to be a benign treatment, sterilization procedures such as 110°C
steam for 15 minutes have been shown to introduce harmful species
into the surfaces of biomaterials. In a paper by Keller et
al., ratios of carbon, oxygen and nitrogen to titanium were
found to increase significantly when analyzed with X-ray Photoelectron
Spectroscopy (XPS). Although hydrogen contents were not specifically
measured, it is likely that these too are elevated after treatment.
In another paper, impurities of fluorine and alkali metals were
found on the implant surface using XPS and Secondary Ion Mass
Spectroscopy (SIMMS). The titanium implant was thought to be protected
from contaminants by keeping it in a titanium box, wrapped in
sterile cloth during the autoclaving procedure. The source of
the contamination was found to be Na2SiF6
which had been used as an additive to the rinsing water in the
final step of the cloth's laundry procedure. The final source
of hydrogen for the implant that must be considered is the fluid
in the joint cavity into which it is placed.
Hydrogen diffuses very quickly in titanium ( Do
ª 5 x 10-3 cm2/sec,
Q = 40.2kJ/mol) and grain boundaries have been shown to
transport the hydrogen much faster than the bulk microstructure:
a fine grained specimen of titanium will absorb over six times
more hydrogen than a large grained specimen. In hcp a
titanium there are two possible positions available for interstitial
atoms: tetrahedral (radius = 0.34Å), and octahedral (radius
= 0.62Å). Hydrogen's radius (0.41Å) prevents it from
occupying the tetrahedral sites. The low solubility of hydrogen
in the a phase is thermodynamically
explained as follows: the decrease in free energy associated with
the introduction of hydrogen in the octahedral sites is offset
by an increase in free energy due to a chemical interaction associated
with the large amount of freedom of the hydrogen atoms to vibrate
within these sites. The titanium-hydrogen phase diagram (Figure
6) shows this decrease in hydrogen solubility in titanium
as temperature is decreased.

The higher solubility of hydrogen in the bcc b
phase of titanium alloys (ª 1
wt% or 35 at% max.) is explained by the close match between the
site radius (now 0.44Å) and hydrogen's atomic radius, eliminating
the free energy increase associated with large atomic hydrogen
vibrations. Additions of interstitial oxygen or nitrogen do not
influence the amount of hydride phase precipitated in titanium
at room temperature. Substitutional aluminum atoms (primarily
present in the a phase of the Ti6Al4V
alloy) inhibit hydride phase formation.
In research conducted by Xiaoxia et al. concerning the
wear behavior of Ti-6Al-4V alloy in an acidic medium, hydrogen
was found to play a key role in determining the extent and nature
of debris generation. In this paper by Xiaoxia et al. the
corrosive wear behavior of Ti-6Al-4V was investigated using a
pin on disk apparatus in 1N H2SO4
at different applied electrochemical potentials. The quantity
of wear debris quadrupled when the wear specimen was polarized
from -0.8 V to -1.4 V as shown in Figure 7 from this reference.
Forced cathodic polarization of the specimen from -1.0 V to -1.4
V correlates to an increase in hydrogen content by a factor of
approximately 2.5. Thus, these authors were able to charge the
titanium alloy with hydrogen by simply applying a negative potential
in an acidic solution.
The hydrogen content of the jagged wear particle polarized to
-1.2V (shown in Figure 8) was approximately 2.3 times higher
than that of the ductile wear particle (Figure 9) and 3.5
times greater than that of the original (unpolarized) alloy specimen,
as inferred from the results of Secondary Ion Mass Spectrometry
(SIMS) (Figure 10). It is crucial to note the difference
in the nature of the debris generated at the two different conditions.
The high-hydrogen content particle is rough, very thin, and exhibits
surface cracking. The low-hydrogen content particle is quite ductile
in appearance and exhibits shear banding. The inert gas fusion
method (the measurement of the amount of hydrogen out-gassed from
the known quantity of metal in an inert gas furnace) was also
used to determine the concentration of hydrogen at the surface
of the disk and within the wear debris. This method found that
polarization of the specimen from -0.48V to -1.2V increased the
hydrogen content by a factor of 4.




Other researchers (T.I. Wu and J-K Wu), in their investigation
of hydrogen as a surface hardening agent for Ti6Al4V, found
that electrolytically charging the alloy in an acidic (or basic)
solution led to a microstructural transformation of the acicular
b phase to the equiaxed a
structure at room temperature. This is significant in that it
emphasizes how fast hydrogen can travel in the alloy and how important
of an effect it can have upon the alloy's microstructure.
For comparative purposes, Figure 11 shows an electron micrograph
of the surface of a Ti6Al4V alloy that has been fretted in air
at a very high load and high number of cycles. The severely damaged
surface exhibits a grooved, jagged, delaminated appearance indicative
of extensive wear.

As described by Wickstrom in Hulbert et al., the body's
fluid is an 'angry' environment that encourages the exchange of
electrolytes and is thus conducive to corrosion. It is oxygenated
at different oxygen partial pressures in different areas. Variation
of the oxygen partial pressure is enhanced by the presence of
various lymphatic cells that secrete enzymes (thus lowering pH)
and absorb oxygen. Figure 12 shows the relationship (measured
in vivo) between pO2
and pH in healthy and diseased synovial fluids. Note how the oxygen
partial pressure begins to become less variable as pH is decreased
and that at normal pH values there is sufficient oxygen in solution
to allow passivation for titanium and cobalt alloys.

Oxygen is brought to all tissues from the bloodstream and ions
in chloride solutions rapidly circulate past any given area. The
pH of the fluid surrounding a recently implanted prosthesis has
been measured to be approximately 5.4, with this value approaching
the 'equilibrium' physiological value of 7.35 within approximately
ten days. However, the pH of the implant site has been shown to
fluctuate to lower values (as low as approximately 4) in cases
of loosening or infection.
A localized concentration of dissolved hydrogen is indicated by
this reduced pH and is likely to be enhanced by large numbers
of crevices in the case of porous coated implants. This increased
hydrogen concentration can certainly alter the microstructure
and, hence, the wear degradation of titanium implants. As we saw
above, titanium has been shown to be highly susceptible to hydrogen
absorption and hydride phase formation in the presence of even
minor quantities of hydrogen.
The placement of the implant into an acidic solution complicates the issue of hydrogen up-take beyond that of gaseous absorption. In order to understand the extent of protection offered by the passive oxide layer discussed above, potential-pH equilibrium diagrams (also referred to as Pourbaix diagrams for their developer, Marcel Pourbaix) can be consulted that indicate regions of phase stability according to environmental conditions. Pourbaix diagrams consider the effect of pH and applied potential upon metallic species. Solid lines on the diagram indicate a threshold of 10-6M concentration of metal ions. Thus, enclosed regions denote the phase (or ionic state) of the metal under consideration when immersed in water. The parallel dashed lines indicate the stability region of water and indicate the concentration of oxygen or hydrogen in solution. At the top line, gaseous oxygen is liberated, governed by the reaction
below the lower line, hydrogen is evolved:
Figure 13 shows the Pourbaix diagram for iron and shows
that iron exhibits two separate regions of corrosion, a region
of passivity, and at the bottom, a region of immunity. The 'Passive'
region indicates that corrosion (though possible) is less likely
than the formation of a protective oxide layer on the metal surface.
The 'Immune' region implies that no corrosion will occur (this
is the region sought for cathodic protection of iron). Finally,
the two 'Corrosion' regions signify the release of iron ions into
solution.

Figure 14 is the Pourbaix diagram for titanium. Some major
differences should be noted: In this figure there is only one
Corrosion region, and three different favored passive oxide phases.
Figure 15 is also a Pourbaix diagram for titanium, however,
in this case, the effect of dissolved hydrogen on the metal's
behavior is considered. The lower regions in this diagram now
indicate that titanium hydride (TiH2)
is the stable phase at all pH values for potentials less than
about -0.8V. Essentially, hydrogen serves to shrink the Corrosion
region (the triangular shaped region in this figure) while elevating
the Immune region.


There is a possibility that piezoelectric potentials generated
by the apatite crystals in the bone surrounding the implant may
contribute to corrosion or hydrogen uptake of the implant. Strain-related
potentials have been recorded for living bone tissue. The potentials
released during typical loading conditions, e.g., walking
or jumping, have not actually been measured, but recorded values
for 0.02 in/min displacement rates (very slow by human loading
standards) are -10 to -20 mV for compressive surface strains of
0.0005 to 0.001. It is important to note that for a given peak
strain, the generated potential increases in a manner roughly
proportional to the square root of the strain rate. When these
values are normalized to a more reasonable displacement rate of
10 cm/s, peak generated voltages may reach approximately -140
mV. Thus, it may be possible for the cyclically stressed femur
to generate potentials on the order of ±100 mV while in contact
with the implant's surface. These potential pulses may help to
encourage depassivation. Also, it has been suggested that these
pulses may interfere with normal bone regeneration and be a factor
in the osteolysis (bone deterioration) frequently observed around
the implant.
Surprisingly, there does not seem to be a significant correlation
between patient factors and failure occurrence. However, an inverse
correlation appears to exist between the amount of use and service
life-span of the implant. Unfortunately, mechanical factors alone
cannot be implicated in the reduction of the service life of the
implant, i.e., failure seems to depend on more than mechanical
factors. As discussed below, many failures appear to be heavily
influenced by the chemical and biological aspects of the implant's
service environment.
Immediately after the operation, the patient is in a brace and
bandaging that prevents any motion of the limb and joint. There
is usually a large amount of swelling and bruising from the surgical
damage to the bone and surrounding tissues. During the patient's
recovery stage, the swelling subsides and the patient is allowed
to apply moderate loads (usually in a swimming pool where buoyancy
prevents full loading) to the implant under the supervision of
a physical therapist. This recovery period is intended to allow
the individual microscopic bone fibers to grow into the porous
coating of the implant. Smooth implants rely on the process of
bone adhesion directly to the metallic surfaces or PMMA cement's
mechanical bonding. Depending on the patient and the time elapsed
since the surgery, there may be a small amount of implant-bone
micromotion, but this amount is difficult to quantify. In the
case of a normal, fully functional implant, normal activities
and, hence, full loading of the implant, can be resumed (with
caution) by the patient once sufficient bone in-growth has occurred.
The design value for the primary applied joint reaction force
at the head and neck of the stem is usually taken as five times
the patient's body weight. This is thought to be a conservative
value, however, many apparently 'gentle' motions of the hip joint
can produce a surprisingly large stress build-up and impulse load
that are often difficult to predict analytically. Dysfunctional
or problematic implant loading can occur when the stem is displaced
too much and becomes separated from the bonding bone fibers (or
bone cement). Canine studies indicate that very little amounts
of micromotion between the stem and bone are required to cause
loosening of cementless implants to occur. In this research by
Bragdon, et al., micromotions of only 40 µm were sufficient
to cause fractures and slippage at the bone-stem interface. Micromotions
on the order of 120 µm caused a layer of dense fibrous tissue
(known as a 'capsule') to form around the implant. This encapsulation
can then cause pain and loosening of the implant and will be discussed
as one of the modes of failure for implants.
There are many modes of failure that depend upon which hip prosthesis
is considered and these are summarized in Table 8. Some
failure modes can present themselves independently of other problems,
whereas modes such as loosening of the stem often occur in conjunction
(or synergistically) with a different mechanism (excessive polyethylene
wear, for example).
The most catastrophic mode of failure is macroscopic fracture
of the stem itself. Due to the high potential for localized corrosive
attack in the area of stress, alternating stresses can fatigue
crack an implant well within the 'safe' fatigue fracture stress
ranges determined under non-corrosive conditions. A subset of
the observed implant fatigue failure modes is known as corrosion
fatigue. Corrosion fatigue is a common failure mechanism for stainless
steel implants, but has been observed in other alloy systems as
well. It initiates predominantly in areas that have been subjected
to design or surface stresses such as notches, machined grooves,
or regions of sharp radii of curvature.
Failure can also occur when the stem becomes loose. It may even
protrude through the outer wall of the femur if it places too
much stress on the inner surface of the cortical (dense) bone
of the femur. The first few weeks after the implantation of the
hip replacement are critical. If the stem is loaded too soon,
any bone-stem bonding that may have formed can be lost. Further
loading will only serve to cause more extensive loosening and
tissue damage. Loosening can also be caused by biological reaction
(most likely to polyethylene or metallic particles surrounding
the implant) where a 'capsule' of fibrous tissue made up of proteins
and inflammatory cells develops around the implant and isolates
it from the bone. This capsule can encourage osteolysis or osteonecrosis
(bone loss or bone death) which further enlarge the cavity intended
to snugly support the implant, leading to even further loosening.
The stem can also loosen when the surrounding bone atrophies (commonly
referred to as demineralization) from disuse. Often, this demineralization
can be caused by a mechanism known to orthopedic researchers as
'Stress Shielding.' Stress shielding can occur when the metallic
stem bears more of the patient's weight than that which is ideal
for bone health and maintenance. Indeed, stress is essential for
minerals to deposit and remain deposited upon the precursor protein
fibers that serve as bone's scaffolding structure. Mother Nature
has taken advantage of the piezoelectric nature of bone crystals.
Normal physiological stresses upon these hydroxylapatite crystals
induce voltages that create a delicate, very local pH range that
encourages precipitation of more hydroxylapatite. Polyethylene
debris released from the articulating surface has also been implicated
in the loosening process owing to its tendency to stimulate bone
dissolving cells (osteoclasts) into action.
| Failure Mode | Causes |
| Macroscopic Fracture | Improper Microstructure
Stress Concentration Improper Design for Load |
| Stress Corrosion Cracking (SCC) | Moderately Corrosive Environment |
| Excessive Ball/Socket Wear | Loss of Passive Film
Poor Material Selection Polymeric Degradation |
| Fretting/Three Body Wear | Generation of Corrosion/Wear Debris Particles on Non-Articulating Surfaces
Debris Migrates to Articulating Surfaces to Enhance Wear/Film Breakdown |
| Loosening/Disassembly | Tissue Reaction to Implant
Excessively High Implant Elastic Modulus Excessive Use/Micromotion Prior to Optimal Bone-Implant Adhesion Improper attachment of CoCrMo ball to Ti stem |
Figure 16 displays the different physical processes that
can occur during sliding wear. Figure 16(a) shows welded
junctions that can form on clean surfaces that lead to material
transfer. Although this has been reported for some implant systems,
it appears to be an infrequent wear mechanism for modular implant
systems. When a hard counterbody repeatedly wears against a ductile
surface, sheet-like wear particles are formed. This mechanism
is shown in Figure 16(b). Polyethylene and titanium, when
coupled with harder CoCrMo alloys often exhibit this form of debris.
Figure 16(c) shows the wear behavior when surface traction
in sliding contact leads to cracking of brittle materials such
as ceramics. Figure 16(d) is very important in that it
shows how the cracking of a brittle surface layer can result in
loose debris particles that later act as abrasives. These wear
debris generation mechanisms can manifest themselves at the ball/acetabular
socket couple or stem surface. After many wear cycles, i.e.,
leg extensions/flexions, portions of the stem can loose their
protective passive oxide film, especially in the presence of abrasive
particles, not only those generated from wear, but also those
present from the manufacturing process. These released particles
will contribute to 'three body' wear as shown in Figure 16.
If a metallic (or ceramic) particle becomes lodged into the relatively
soft polyethylene liner, extensive depassivating scratches can
form on the CoCrMo head. Chromia particles released in this manner
may attack the lower portion of the stem, although the results
of most research indicates that extensive CoCrMo head surface
wear is uncommon and that very little Co, Cr, Mo, or Ni (a minor
constituent of this alloy) are detected in wear debris.

Figure 16
Extensive research has also been conducted on the concentration
of metallic ions near the implant in an attempt to demonstrate
the extent of simple corrosion of the implant. One interesting
finding obtained from these metallic ion concentration measurements
is that ions are often found in proportions that differ significantly
from those expected from alloy compositions. Thus, leaching and/or
corrosion of the implant is suspected. interestingly, these constituents
are found in above normal ionic concentrations in regions as removed
from the implant as the liver when any burnishing or fretting
damage of the CoCrMo head is observed. This information implies
that the elemental and ionic species of this alloy are much more
soluble in physiological tissue and fluid than those of Ti6Al4V.
This observation can be explained by the extremely high thermodynamic
instability of the titanium ion in the body. Concentrations of
aluminum and vanadium ions vary considerably and are found to
be elevated in certain regions around the implant as shown in
Table 9. This table shows that ion concentration not only
varies according to patient (the reason for the large scatter)
but also by the type of ion and the location in which it is measured.
| Ion Concentrations (µg/l) | |||||||
| Tissue Type | Ti | Al | V | ||||
| Synovial Fluid | 556±882
(13±22) | 654±743
(109±158) | 62±95
(5±1) | ||||
| Capsule | 1540±1238
(723±1217) | 2053±1064
(951±586) | 288±133
(122±123) | ||||
| Fibrous Membrane | 20813±26467
(N/A) | 10581±9764
(N/A) | 1027±702
(N/A) | ||||
| Blood | 67±62
(17±60) | 218±233
(13±4) | 23±31
(6±4) | ||||
| Co | Cr | Mo | Ni | ||||
| Synovial Fluid | 588±427
(5±3) | 385±232
(3±4) | 58±53
(21±8) | 32±16
(5±2) | |||
| Capsule | 821±451
(25±17) | 3329±2890
(133±63) | 447±247
(17±8) | 5789±2535
(3996±6237) | |||
| Fibrous Membrane | 2229±1583
(N/A) | 12554±8055
(N/A) | 1524±1399
(N/A) | 13234±10074
(N/A) | |||
| Blood | 20±25
(0.1-1.2) | 110±150
(2-6) | 10±4
(0.5-1.8) | 29±29
(2.9-7) | |||
Debris can be generated at many locations on the implant surface
either by corrosive or bone-implant micromotion induced wear mechanisms
collectively termed fretting. This debris can then migrate through
the fluid filled environment to other surfaces and disrupt the
protective passive film. Once the passive film is disrupted in
a tribochemical environment, the fretting-wear process often becomes
synergistic. A corrosive environment inhibits repassivation after
the oxide film has been mechanically removed. Fretting of the
exposed surface leads to more wear particles, which in turn cause
further removal of the protective film, causing even more degradation
of the surface. Figure 17 shows how titanium responds to
passive film disruption in a 0.1 N NaCl solution. Peaks
on this plot of potential vs. time correspond to oxide film growth
while drops in potential are caused by mechanical disruption of
the film. Note how the film restores itself when the fretting
motion is stopped (at 70 minutes).

The following specific origins of metallic debris have
been postulated:
Wear can occur at the ball and socket surfaces of the device that
are exposed to frequent relative motion. Older implant designs
that used titanium for the head (ball) of the stem (and that had
not been surface hardened) were prone to rapid and extensive wear
at the articulative polyethylene socket-titanium head interface.
More recent designs have incorporated a head of different, harder
material, such as CoCr alloys or Al2O3,
to increase the wear resistance of the implant surface. One drawback
observed with the use of the harder head materials is accelerated
wear of the polyethylene liner of the acetabular (hip) socket.
Adverse biological reactions at the surface and in the vicinity
of the implant are now believed to be caused by sub-micron polyethylene
debris particles.
As shown in Figure 18, there is a high potential for fretting
at the press-fit taper joint between the neck of the stem and
the ball. This opportunity for fretting wear is facilitated by
the abutment of two different materials of different hardness
at the points of contact of the couple. A common taper combination
is 5° 59' 57" for the stem cone and 5° 53' 32"
for the bore of the head. Tolerances for medical tapers are up
to 8 times looser than those for the automotive and machine tool
industries. Taper joints exposed to both actual and simulated
implant loading conditions have shown signs of both fretting wear
and corrosion. Collier and other researchers have also noted this
phenomenon for cobalt alloy balls on cobalt alloy stems. When
in contact, two metals (e.g., CoCrMo and Ti-6Al-4V) of
different galvanic potential will establish an anodic-cathodic
reaction that can exacerbate the corrosion process, especially
in the presence of a crevice (as shown in Figure 18). Although
this theory is strongly advocated by Collier, a more recent study
(sponsored by the orthopedic company Zimmer, Inc.) that examined
the corrosion behavior of nitrided Ti6Al4V/CoCrMo alloy taper
joints in saline solution under a 700 lb head load for one million
cycles found that only crevice corrosion was evident. No micromotion
induced damage to the taper joint was observed. The authors conclude
that the nitriding process prevented surface damage/depassivation.

Motion of the stem portion of the implant relative to the surrounding
compact or cancellous (spongy) femoral bone could be implicated
in removing the passive and protective oxide film of the titanium
stem. When high resolution surface analysis techniques (such as
atomic force microscopy or X-ray microanalysis) are applied to
Ti6Al4V implant materials, needle shaped oxide growths are found
on the surface (Figure 19). Apatite, the main hard mineral
component of bone, possesses a hardness of approximately 35 on
the Rockwell 'C' scale or 5 on the Mohs hardness scale where diamond
is 10 and talc is 1. It is possible that this hard component of
the bone could selectively remove portions of the oxide and underlying
metal. Retrieved implants often exhibit 'shiny' regions on their
normally matte surfaces where the implant had been in direct contact
with the inside of the femur.

The sintered porous mesh or bead surface layers used on some stems possess elastic properties that are much lower (due to their lower strength and smaller cross-sectional areas) than those of the substrate material. Hooke's law for a two component composite material (in this case, the implant, consisting of the stem and outer mesh) states that
These loose regions may contribute to depassivation of the underlying
surface when they are cyclically strained during the implant's
service. Evidence for this mechanism of particle generation is
difficult to establish since the entire process would be occurring
beneath the sintered pad or beaded surface and has not yet been
substantiated in the current literature. The increased surface
area of these porous regions coupled with an increased probability
of finding defects (such as hydrides or crevices) at the implant's
surface defects tends to strongly implicate the porous mesh in
the debris generation process. Indeed, many researchers have found
that when retrieved porous coated implants are examined, the porous
layer had delaminated from the surface and that metallic debris
and associated biological inflammation were prevalent in the tissue
next to the coating.
This mechanism of debris formation is only operative in implant
systems that are bonded to the femur with polymethylmethacrylate
bone cement (not the implants considered by this research). The cement relies upon mechanical bonding in order to adhere
to the interior of the femur, but relies primarily upon chemical
bonding to secure the stem within the femur. If the stem manages
to loosen from the (primarily chemical) bond of the cement, it
will rub and wear against the fairly brittle cement sleeve. Burnished
areas are often noted on implants recovered from loosened cement
capsules that are indicative of macroscopic wear.
The American Society for Testing and Materials (ASTM) Committee F-4 on Medical and Surgical Materials and Devices has established a task force to develop methods for the characterization of particulate debris. Once these methods have been established, researchers will be able to communicate their findings more uniformly and accurately. To date, very few sources have reported on the morphological aspects of the particulate debris released from Ti6Al4V modular hip joint prostheses.
Next Chapters (Chapters 6-8) Previous Chapter (Chapter 4) Table of Contents
Some files on this site are in .pdf format. To read these you will need the Adobe Acrobat Reader.
|